A cold silicon strip telescope for time-of-flight determination at −55°C has been tested using a hadron beam at Saturne II. Saclay. We present performance tests of the telescope, the time-of-flight distributions between pairs of silicon hodoscopes and their dependence on temperature. A detailed description of the apparatus, the refrigerator system and the calibration procedures is also reported. A linear relationships between the time resolution and the temperature of the apparatus is measured. An improvement of a factor 2 in the time-of-flight resolution is observed when the temperature decreases from +20°C to −55°C

Performance of a time-of-flight silicon strip telescope in the temperature range 20°C to -55°C

GRIMANI, CATIA;
1995

Abstract

A cold silicon strip telescope for time-of-flight determination at −55°C has been tested using a hadron beam at Saturne II. Saclay. We present performance tests of the telescope, the time-of-flight distributions between pairs of silicon hodoscopes and their dependence on temperature. A detailed description of the apparatus, the refrigerator system and the calibration procedures is also reported. A linear relationships between the time resolution and the temperature of the apparatus is measured. An improvement of a factor 2 in the time-of-flight resolution is observed when the temperature decreases from +20°C to −55°C
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11576/2513797
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 6
social impact